TELESTO Nitrogen Generator

PSA Technology

Reference

TELESTO.[Flow rate]/0 (Without compressor) TELESTO.[Flow rate]/1 (With compressor)

With 4 models, 4 L/min to 15 L/min, the TELESTO Nitrogen range provide a nitrogen solution for ELSD / CORONA instrument.
The TELESTO is available without or with integrated oil free air compressor using the Energy Saving Technology mode (EST). It is extremely compact and quiet in operation.
The generator is controlled using the latest in color touch screen technology to display the process in real time, inlet / outlet pressure, N2 Flow and status of the system with remote access for diagnostic control.
Utilizing Advanced technology PSA of separating Nitrogen and Oxygen to produce high quality Nitrogen gas on demand 24/7.

Advantages product

INCREASED LABORATORY EFFICIENCY

A constant, uninterrupted gas supply of guaranteed purity
eliminates interruptions of analyses to change cylinders and
reduces the amount of instrument re-calibrations required.

IMPROVED ECONOMY

Pure nitrogen gas produced as standard.

IMPROVED SAFETY

Nitrogen produced at low pressure and ambient temperature
removes the need for high pressure cylinders.

SECURITY OF SUPPLY

Integral oil free air compressor as an option guarantees
continuous gas supply, independent of in house compressed
air supply.

SIMPLE INSTALLATION

Gas generators can be installed in the laboratory, on or
under a bench, eliminating the need for long gas lines from
cylinders secured elsewhere.

Characteristics

  • Intelligent color touch screen to provide simple and user friendly management of all functions of the unit
  • Fully regenerative PSA technology
  • Without or with integral oil free air compressor
  • Quiet thanks to the Soundproofed compressor box and anti-vibration features
  • Auto start
  • Audible alarm display and Visual maintenance indication with history log
  • Outlet flow indicator
  • Energy Saving Technology (EST):
    - Switch off the compressor when gas supply is not required
    - Allowing low energy consumption and increasing durability of the compressor
  • Remote PC monitoring and diagnostic analysis via USB/ RS232 to interface the unit with customer’s PC software (allow to carry out checks and maintenance effectively, only via a remote connection)
  • Compact design
  • Fit with wheels
  • Specifications

    Models
    TELESTO.4
    TELESTO.6
    TELESTO.8
    TELESTO.15
    GENERALS INFORMATIONS
    Compressor
    0 (Without)
    1 (With)
    0 (Without)
    1 (With)
    0 (Without)
    1 (With)
    0 (Without)
    1 (With)
    Max Nitrogen Flow rate
    4 L/min
    6 L/min
    8 L/min
    15 L/min
    Max Nitrogen outlet pressure
    6 bar (87 psi)
    N2 purity
    99.5%
    Particles
    0.01 μ
    Inlet Air flow required @ 8 bar (116 psi) max
    14 L/min
    N/A
    21 L/min
    N/A
    28 L/min
    N/A
    52 L/min
    N/A
    Min Inlet Air quality
    ISO8573 –
    1:2010 1.4.1
    N/A
    ISO8573 –
    1:2010 1.4.1
    N/A
    ISO8573 –
    1:2010 1.4.1
    N/A
    ISO8573 –
    1:2010 1.4.1
    N/A
    Touch screen
    Touch screen (operating parameters, system status, alarms) with LED indicators(Power on/off; ready or errors)
    Max Relative Humidity
    80% RH @ 25°C (77°F)
    Temperature range
    From 5 – 25°C (41 – 77°F))
    Max Altitude
    3000 Meters
    Noise level @ 1M
    25 dB (A)
    50 dB (A)
    25 dB (A)
    50 dB (A)
    25 dB (A)
    52 dB (A)
    25 dB (A)
    52 dB (A)
    Dimensions
    (W x H x D)
    47.5 x 66.2 x 65 cm
    (18.7″ x 26″ x 25.6″)
    47.5 x 66.2 x 79 cm
    (18.7″ x 26″ x 31″)
    Weight (kg/lbs)
    54 / 119
    58 / 127
    56 / 122
    60 / 132
    88 / 194
    108 / 238
    91 / 200
    118 / 260
    Power consumption
    150 VA
    375 VA
    150 VA
    375 VA
    150 VA
    850 VA
    150 VA
    1700 VA
    Electrical supply
    220 – 240 V ac / 1 ph / 50 – 60 Hz
    110 – 120 V ac / 1 ph / 50 – 60 Hz
    Fuse
    10A
    CONNECTIONS
    Nitrogen outlet
    G 1/4” (BSP) Female
    Air Inlet
    G 1/4”(BSP)Female
    N/A
    G 1/4”(BSP)Female
    N/A
    G 1/4”(BSP)Female
    N/A
    G 1/4”(BSP)Female
    N/A
    COMMUNICATION
    USB/PC Control
    In series

    Download documentation

    Having all the necessary information to successfully complete your project is essential, which is why we are providing you with these documents.

    Please feel free to contact us with any questions.

    This site is registered on wpml.org as a development site. Switch to a production site key to remove this banner.